The Journal of the American Dental Association
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Am Dent Assoc, Vol 130, No 9, 1354-1359.
© 1999 American Dental Association

This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by MUZZIN, K. B.
Right arrow Articles by BERRY, C. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by MUZZIN, K. B.
Right arrow Articles by BERRY, C. W.
Related Collections
Right arrow Dental Equipment/Instruments

ADVANCES IN DENTAL PRODUCTS

JADA Continuing Education

ASSESSING THE CLINICAL EFFECTIVENESS OF AN AEROSOL REDUCTION DEVICE FOR THE AIR POLISHER



KATHLEEN B. MUZZIN, R.D.H., M.S., TRACY B. KING, R.D.H., B.S. and CHARLES W. BERRY, PH.D.

Background. Because of the growing concern of the infectious nature of aerosols produced by air polishers, the Centers for Disease Control and Prevention and the American Dental Association have made recommendations to minimize dental aerosols. In addition, a device was developed in 1997 to reduce the aerosol generated during air polishing.

Methods. The authors conducted a blinded clinical study to determine the effectiveness of an aerosol reduction device during air polishing. All 30 subjects underwent two minutes of air polishing, one with the aerosol reduction device on one side of the mouth and one without the device on the opposite side. Microbial samples were collected on blood agar plates positioned 12 inches from the subject’s mouth and from one plate attached to the dental hygienist’s face mask. The sample plates were then inverted and incubated for three days at 37 C.

Results. Using the paired t-test, the authors found that the air polisher without the aerosol reduction device generated a significantly greater mean quantity of colony-forming units, or CFUs, 12 inches from the subject’s mouth (mean ± standard deviation, 148.00 ± 145.00 CFUs) than the air polisher with the aerosol reduction device (20.10 ± 53.90 CFUs) (P < .0001). In addition, the air polisher without the aerosol reduction device generated a significantly greater number of CFUs on the face mask plate (40.90 ± 33.80 CFUs) than the air polisher with the aerosol reduction device (8.80 ± 15.10 CFUs) (P < .0001).

Conclusion. The data suggest that the aerosol reduction device is effective in reducing the number of microorganisms generated during air polishing.

Clinical Implications. Air polishing with the aerosol reduction device may decrease the potential for disease transmission.




This article has been cited by other articles:


Home page
Journal of the American Dental AssociationHome page
S. K. HARREL and J. MOLINARI
Aerosols and splatter in dentistry: A brief review of the literature and infection control implications
J Am Dent Assoc, April 1, 2004; 135(4): 429 - 437.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright©1995-1999 American Dental Association (ADA).
Reproduction or republication strictly prohibited without prior written permission of ADA.