J Am Dent Assoc, Vol 138, No 2, 196-201.
© 2007 American Dental Association

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Di Fiore, P. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Di Fiore, P. M.
Related Collections
Right arrow Dental Equipment/Instruments

CLINICAL PRACTICE

JADA Continuing Education

A dozen ways to prevent nickel-titanium rotary instrument fracture



Peter M. Di Fiore, DDS, MS


   ABSTRACT
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
Background and Overview. With the increased use of nickel-titanium (NiTi) rotary instruments for root canal preparation in endodontics, instrument fracture has become more prevalent. Extensive research has been conducted on the physical properties and mechanical characteristics of NiTi rotary instruments, as well as the factors that can contribute to instrument failure. NiTi rotary instruments are subjected to torque and are susceptible to cyclic fatigue, which are the main causes of instrument fracture. However, with an understanding of how these instruments function in preparing root canals and by applying ways to reduce torque-generated metal fatigue, clinicians can use the instruments safely in clinical practice.

Results. The author presents 12 measures that clinicians can take to prevent NiTi rotary instrument fracture and discusses them in detail.

Clinical Implications. NiTi rotary instrument fracture complicates the progress, and compromises the prognosis of endodontic treatment. However, when clinicians take appropriate measures, rotary instrument fractures can be prevented.

Key Words: Nickel-titanium; rotary; instrument; fracture; prevention

Abbreviations: NiTi: Nickel-titanium • SEM: Scanning electron microscopic

During the past 15 years, nickel-titanium (NiTi) rotary instruments have become a part of the standard armamentarium in endodontics. They are used extensively by generalists and specialists to facilitate the cleaning and shaping of root canals,1 and it appears that with the increased application of these instruments in contemporary endodontic practice, fractures have become more prevalent.2,3

Fractured instruments are a definite hindrance to the goals of cleaning, shaping and filling root canals,4,5 and they may adversely affect the outcome of endodontic treatment.2,68 Techniques for removing fractured instrument fragments from root canals have been described in the dental literature.9,10 However, removal of fragments may be impossible or impractical, especially when they are small and located in the apical portion of narrow curved root canals or when repeated attempts at removal could result in excessive enlargement of the canal or perforation of the root.7,8

Since instrument fracture complicates and compromises endodontic treatment and prognosis,2,68 clinicians must be constantly aware of the possibility of rotary instrument fracture and take every precaution to avert this mishap. The purpose of this article is to present measures that clinical operators can take to reduce the risk of NiTi rotary instrument failure and prevent fracture during root canal preparation.


   ROTARY INSTRUMENT ASSESSMENT
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
NiTi. NiTi is a superelastic shape-memory metallic alloy that, when flexed, undergoes a martensitic transformation from its original austenitic structure and, if stressed beyond its elastic limit, will rupture. During root canal preparation, NiTi rotary instruments are subjected to cyclic fatigue, which can lead to distortion and fracture, especially when the instruments are flexed severely.11 Extensive investigations of the physical properties of NiTi rotary files under dynamic testing procedures have shown that torsional stress and cyclic fatigue are the main causes of instrument fracture.1114 Observations of the fractured surfaces of NiTi rotary instruments under scanning electron microscopic (SEM) examination have revealed the presence of peripheral serrations, dimples and craters that are characteristic of ductile-type fractures.1315 Therefore, NiTi rotary instruments should not be subjected to excessive torsional and bending stress during operation.

Size, taper and cutting flute depths are important factors that affect the torsional and bending properties of rotary instruments.

Instrument design. NiTi rotary instruments are available in a variety of types, with different functional features that affect the manner in which they engage and cut dentin. The structural characteristics and mechanical designs of these instruments have a definite influence on their susceptibility to fracture.1519 In particular, the size, taper and cutting flute depths are important factors that affect the torsional and bending properties of rotary instruments.1519

Size. A comparative study of the fatigue resistance of NiTi rotary instruments of different sizes and flute designs revealed that large instruments were highly susceptible to fatigue failure.15 Research has demonstrated that as an instrument’s cross-sectional diameter increases, it becomes less resistant to cyclic fatigue.18 After performing dynamic stress tests on various NiTi rotary instruments, investigators found that as an instrument’s size and taper increase, the torque generated during rotation increases and the fracture time decreases.12,13

Taper. In cyclic fatigue-to-fracture tests of different NiTi rotary instruments, researchers found that 0.06 taper instruments had less resistance to fracture than did 0.04 taper instruments.19 Shen and colleagues20 compared the types of failures that occurred with NiTi rotary instruments of various geometric designs and reported that a very high percentage (21 percent) of the instruments that fractured had progressively larger tapers and a much lower percentage (7 percent) had consistently even tapers. They also noted that failures for progressively tapered instruments tended to be fractures, whereas for evenly tapered instruments, failures tended to be unwinding deformations.20 In this regard, instruments that show unwinding as a failure characteristic may be safer for use than those that fracture spontaneously. Additionally, Guilford and colleagues,21 in comparing the torque required to fracture different types of rotary instruments, found that progressively tapered instruments failed rapidly with little rotation.

Cutting flute depth. Instruments with deep cutting flutes and progressively larger variable tapers have rapidly changing cross-sectional diameters along the entire length of their shafts. These instruments develop high torque levels that make them more prone to metal fatigue and fracture. However, instruments that have shallow cutting flutes, evenly tapered shafts and consistently shaped cross-sectional areas are more resistant to fracture. This is because the torsional and bending stresses that develop during use are distributed uniformly along these instruments’ entire length.1921 Therefore, practitioners should be completely familiar with the mechanical features and working limitations of rotary instruments and select those that are less prone to fracture.

Instrument use. Microstructure and surface analysis of unused NiTi rotary instruments revealed that there were distortions in the lattice structure of the alloy, variations in the microhardness of the metal, and machining and milling marks as well as metal strips and microcracks on their surfaces.16,22,23 It is important for clinicians to realize that these pre-existing conditions associated with the manufacturing process may contribute to the propagation of instrument fractures during use.24,25 Cyclic fatigue and torsional testing procedures that measured rotation time and torque level at fracture have demonstrated that used rotary instruments are significantly more susceptible to fracture than are new ones.26,27 These findings are further supported by SEM observations of used instruments that revealed signs of deterioration, including surface cracks that can progress to fractures with further use.2225,28 Sotokawa29 found that by applying a systematic schedule for the disposal of endodontic instruments, the incidence of fracture can be reduced.

Practitioners should enlarge root canal orifices before introducing nickel-titanium rotary instruments into the canal.

Therefore, it is advisable and prudent to dispose of all instruments after they have been used for a specific number of clinical cases, rather than wait for deformations and distortions to appear. Manufacturers recommend that rotary instruments be discarded after they have been used for one clinical case.


   CANAL CURVATURE ASSESSMENT
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
The fracture potential of an instrument rotating in a curved canal becomes greater as the angle of curvature increases and the radius of curvature decreases.11,13,14 Zelada and colleagues30 and Martin and colleagues31 reported that during the preparation of root canals in extracted molar teeth, all instrument fractures occurred in severely curved canals with angles of curvature greater than 30 degrees. A careful preoperative radiographic examination with fine hand instruments in the canals will reveal the presence and acuity of root canal curvatures. Therefore, when curvatures are present, the operator should be wary of the possibility of a fracture and proceed cautiously during root canal preparation.


   ACCESS PREPARATION
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
In the internal configuration of an adequate endodontic access preparation, the entrances to the root canal orifices are not obstructed by the presence of excessive dentinal bulk or restorative material, and there is an unimpeded direction from which instruments can approach the apical portion of the root canal or the point of initial root canal curvature.32 When instruments can negotiate root canals easily, bending and flexing stresses are lessened and the potential for fracture is reduced. Endodontic access becomes even more crucial for avoiding an instrument fracture when teeth are difficult to reach because of limited mouth opening.33 Therefore, practitioners should make adequate access preparation a priority, as an important first step in avoiding rotary instrument fracture.


   CANAL ORIFICE ENLARGEMENT
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
The enlargement of root canal orifices facilitates the negotiation and instrumentation of the apical part of root canals, especially in curved canals of multi-rooted teeth.34,35 Leeb34 used maxillary and mandibular extracted molars with curved roots to demonstrate that after the canal orifices were enlarged, instruments more easily penetrated the canals. The canal orifice can be enlarged effectively by the sequential use of nos. 4 and 2 low-speed, long-shank round burs followed by nos. 4 and 3 Gates Glidden drills.34,35 These instruments, used carefully, can efficiently create a 2- to 4-millimeter oval-shaped funnel that serves as an accessible unobstructed entrance and guides rotary instruments into the root canal without causing excessive bending or binding, which could lead to metal fatigue. Therefore, practitioners should enlarge orifices before introducing NiTi rotary instruments into the root canal.


   MANUAL INSTRUMENTATION
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
Hand instruments can create a smooth, open passageway for rotary instruments to follow as they progress to the apical terminus. Three studies have demonstrated that manual root canal instrumentation with fine stainless steel hand instruments, used in a step-back manner before rotary instruments were used, significantly reduced the incidence of rotary instrument fracture during the preparation of curved canals.3638 Roland and colleagues36 and Patino and colleagues37 used fine hand instruments to enlarge curved root canals in extracted molars manually, and Berutti and colleagues38 manually enlarged standardized curved canals in resin blocks to a size 20 hand instrument, creating a glide path for rotary instruments. All of these studies showed that manual enlargement of root canals with fine hand instruments significantly reduced the failure rate of rotary instruments.3638 Therefore, practitioners should use rotary instruments only after root canals have been negotiated and enlarged with fine hand instruments.


   ROTATIONAL SPEED AND TORQUE CONTROL
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
Electric motors have been developed to control both rotational speed and torque during root canal instrumentation so that when the torque on an instrument, rotating at a constant speed, reaches a preset level, the motor automatically reverses its rotational direction and allows the file to be withdrawn before it locks and fractures in the root canal.39 Gabel and colleagues40 investigated the influence of rotational speed on the failure of NiTi rotary instruments for the preparation of root canals in extracted molar teeth and found that instrument distortion and fracture were four times more likely to occur at higher rotational speeds (333 rotations per minute) than at lower rotational speeds (167 rpm). Gambarini41 found that instruments used in low-torque motors (< 1 Newton per centimeter) were more resistant to fracture than those used in high-torque motors (> 3 N/cm). Therefore, practitioners should use electric motors set at low rotational speeds and low torque levels during root canal preparation.


   CROWN-DOWN TECHNIQUE
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
In the crown-down technique, larger instruments are used first in the coronal aspect of the canal, followed sequentially by smaller instruments in the apical aspect of the canal. The advantages of this technique are that it removes infected coronal dentin and obstructions before apical preparation and enlarges canals incrementally. In a study by Blum and colleagues42 in which the root canals of mandibular incisors were prepared with NiTi rotary instruments using either a step-back or crown-down technique, the researchers found that less vertical force and torque were created with the crown-down technique and that instrument tips had less contact with dentin and less stress during the early phases of instrumentation. This is important, since test fractures performed on NiTi rotary instruments have demonstrated that fractures tend to occur close to the tip.43 Studies of the failure of NiTi rotary instruments used to prepare root canals in extracted molar teeth found that a greater number of fractures and distortions occurred with sizes 20 and 25, and that most instruments fractured within 1 to 3 mm from the tip.30,40,44 Additionally, in two separate investigations of failure of rotary instruments that were used with the crown-down technique to prepare a total of 210 curved root canals in extracted teeth, researchers found that no fractures occurred and only two instruments became deformed.45,46 Therefore, practitioners should apply the crown-down technique as a standard operational procedure with rotary instruments.


   IRRIGATION AND LUBRICATION
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
Irrigation and lubrication are essential for accomplishing adequate débridement of root canals. SEM studies of the efficacy of root canal cleaning have demonstrated that dentinal debris generated during instrumentation becomes packed in root canals and that irrigation is necessary for its removal.4749 A preparation of urea peroxide and ethylenediamine tetraacetic acid as a lubricant50 is a combination commonly used for root canal preparation.1 Irrigants and lubricants reduce root canal clogging, frictional resistance and mechanical overloading, thereby decreasing the torsional stresses placed on rotating instruments. Therefore, during root canal preparation, practitioners should lubricate instruments generously and irrigate canals copiously.


   ROTARY INSTRUMENT MANIPULATION
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
The manner in which NiTi rotary files are manipulated for preparing root canals is extremely important. It has been shown that a cyclic axial motion applied to rotary instruments during operation was significant in preventing premature fracture.51 Also, a pecking or pumping motion, which lowers apical forces during root canal preparation, has been advocated by researchers as an important way to prevent instrument binding and torque-generated cyclic fatigue.1214

Li and colleagues14 tested the cyclic fatigue of NiTi rotary instruments under static and dynamic pecking motion conditions and found that as the pecking distance increased, the fracture time increased, suggesting that this type of instrument manipulation is critical for preventing rotary instrument fracture. Yared and colleagues45 found that when a slight apical pumping motion was used to reduce the development of excessive torque during the root canal preparation of extracted molar teeth, no instruments fractured. Therefore, one may conclude that practitioners should use a pecking or pumping movement when manipulating rotary instruments.


   OPERATOR PROFICIENCY
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
Studies have demonstrated that higher rates of NiTi rotary instrument fracture occur with inexperienced operators than with experienced ones.52,53 Rotary instruments tend to thread and screw into root canals, which subjects them to high levels of torque as they bind and lock in the canal.39,52,53 In addition, instrument locking may be enhanced when the root canal preparation begins to acquire the shape and taper of larger instruments as they extend deeper into the canal, creating a taper-lock effect.39,52,53 This is a valid concern; Schrader and Peters54 found that using NiTi rotary instruments with different tapers reduced canal contact areas and instrument fatigue–related failures during root canal preparation in extracted teeth. The operator’s ability to sense and resist these binding and locking tendencies is a skill that can be obtained only with experience. Yared and colleagues,39,52,53 in several extensive investigations, showed that preclinical training in the use of NiTi rotary instruments for the preparation of root canals in extracted molar teeth was crucial for avoiding instrument fracture. Therefore, inexperienced operators should engage in preclinical training exercises as learning experiences before using these instruments on patients, then proceed carefully in clinical practice as they gain experience.


   SUMMARY AND CONCLUSION
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 
There are several measures that practitioners can take to prevent NiTi rotary instrument fracture during root canal preparation:

– avoid subjecting NiTi rotary instruments to excessive stress;
– use instruments that are less prone to fracture;
– follow an instrument use protocol;
– assess root canal curvatures radiographically and instrument them carefully;
– ensure that the endodontic access preparation is adequate;
– open orifices before negotiating canals;
enlarge root canals with fine hand instruments;
– set rotational speed and torque at low levels;
– use the crown-down technique;
– irrigate and lubricate root canals during preparation;
– manipulate rotary instruments with a pecking or pumping motion;
– if inexperienced, engage in preclinical training in the use of rotary instruments.

Instrument fracture is a serious iatrogenic mishap that can complicate and compromise endodontic treatment. It therefore is imperative that clinicians using these instruments in practice apply all appropriate measures to reduce the risk of fracture. Recently published laboratory and clinical assessment studies have shown that when operators are aware of the possibility of instrument fractures and take measures to avoid them, the incidence of fracture can be as low as four per 1,000.55,56


   FOOTNOTES
 

Dr. Di Fiore is an associate professor of endodontics and the director, Predoctoral Endodontics, New York University, College of Dentistry, Department of Endodontics, 345 E. 24th St., New York, N.Y. 10010. Address reprint requests to Dr. Di Fiore.


   REFERENCES
 TOP
 ABSTRACT
 ROTARY INSTRUMENT ASSESSMENT
 CANAL CURVATURE ASSESSMENT
 ACCESS PREPARATION
 CANAL ORIFICE ENLARGEMENT
 MANUAL INSTRUMENTATION
 ROTATIONAL SPEED AND TORQUE...
 CROWN-DOWN TECHNIQUE
 IRRIGATION AND LUBRICATION
 ROTARY INSTRUMENT MANIPULATION
 OPERATOR PROFICIENCY
 SUMMARY AND CONCLUSION
 REFERENCES
 

  1. Averbach RE, Kleier DJ. Endodontics in the 21st century: the rotary revolution. Compend Contin Educ Dent 2001;22(1):27–34.[Medline]

  2. Spili P, Parashos P, Messer HH. The impact of instrument fracture on outcome of endodontic treatment. J Endod 2005;31(12):845–50.[Medline]

  3. Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod 2004;30(10):722–5.[Medline]

  4. Schilder H. Cleaning and shaping the root canal. Dent Clin North Am 1974;18(2):269–96.[Medline]

  5. Schilder H. Filling root canals in three dimensions. Dent Clin North Am 1967;11:723–44.

  6. Crump MC, Natkin E. Relationship of broken root canal instruments to endodontic case prognosis: a clinical investigation. JADA 1970;80(6):1341–7.

  7. Hulsmann M, Schinkel I. Influence of several factors on the success or failure of removal of fractured instruments from root canals. Endod Dent Traumatol 1999;15(6):252–8.[Medline]

  8. Souter NJ, Messer HH. Complications associated with fractured file removal using an ultrasonic technique. J Endod 2005;31(6):450–2.[Medline]

  9. Hulsmann M. Removal of fractured instruments using a combined automated/ultrasonic technique. J Endod 1994;20(3):144–6.[Medline]

  10. Ward JR, Parashos P, Messer HH. Evaluation of an ultrasonic technique to remove fractured rotary nickel-titanium endodontic instruments from root canals: an experimental study. J Endod 2003;29(11):756–63.[Medline]

  11. Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997;23(2):77–85.[Medline]

  12. Sattapan B, Palamara JE, Messer HH. Torque during canal instrumentation using rotary nickel-titanium files. J Endod 2000 26(3):156–60.[Medline]

  13. Haikel Y, Serfaty R, Bateman G, Singer B, Allemann C. Dynamic and cyclic fatigue of engine driven rotary nickel-titanium endodontic instruments. J Endod 1999;25(6):434–40.[Medline]

  14. Li UM, Lee BS, Shih CT, Lan WH, Lin CP. Cyclic fatigue on endodontic nickel titanium rotary instruments: static and dynamic tests. J Endod 2002;28(6):448–51.[Medline]

  15. Chaves Craveiro de Melo M, Guiomar de Azevedo Bahia M, Lopes Buono VT. Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments. J Endod 2002;28(11):765–9.[Medline]

  16. Kuhn G, Tavernier B, Jordan L. Influence of structure on nickel-titanium endodontic instruments failure. J Endod 2001;27(8):516–20.[Medline]

  17. Xu X, Eng M, Zheng Y, Eng D. Comparative study of torsional and bending properties for six models of nickel-titanium root canal instruments with different cross-sections. J Endod 2006;32(4):372–5.[Medline]

  18. Ullmann CJ, Peters OA. Effects of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments. J Endod 2005;31(3):183–6.[Medline]

  19. Yao JH, Schwartz SA, Beeson TJ. Cyclic fatigue of three types of rotary nickel-titanium files in a dynamic model. J Endod 2006;32(1): 55–7.[Medline]

  20. Shen Y, Cheung GS, Bian Z, Peng B. Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 2006;32(1):61–5.[Medline]

  21. Guilford WL, Lemons JE, Eleazer PD. A comparision of torque required to fracture rotary files with tips bound in simulated curved canal. J Endod 2005;31(6):468–70.[Medline]

  22. Tripi TR, Bonaccorso A, Tripi V, Condorelli GG, Rapisarda E. Defects in GT rotary instruments after use: an SEM study. J Endod 2001;27(12):782–5.[Medline]

  23. Martins RC, Bahia MG, Buono VT. Surface analysis of ProFile instruments by scanning election microscopy and X-ray energy-dispersive spectroscopy: a preliminary study. Int Endod J 2002;35(10): 848–53.[Medline]

  24. Alapati SB, Brantley WA, Svec TA, Powers JM, Nusstein JM, Daehn GS. SEM observations of nickel-titanium rotary endodontic instruments that fractured during clinical use. J Endod 2005;31(1):40–3.[Medline]

  25. Alapati SB, Brantley WA, Svec TA, Powers JM, Mitchell J. Scanning electron microscope observation of new and used nickel-titanium rotary files. J Endod 2003;29(10):667–9.[Medline]

  26. Gambarini G. Cyclic fatigue of ProFile rotary instruments after prolonged clinical use. Int Endod J 2001;34(5):386–9.[Medline]

  27. Yared G. In vitro study of the torsional properties of new and used ProFile nickel titanium rotary files. J Endod 2004;30(6):410–2.[Medline]

  28. Svec T, Powers JM. The deterioration of rotary nickel-titanium files under controlled conditions. J Endod 2002;28(2):105–7.[Medline]

  29. Sotokawa T. A systemic approach to preventing intracanal breakage of endodontic files. Endod Dent Traumatol 1990;6(2):60–2.[Medline]

  30. Zelada G, Varela P, Martin B, Bahillo JG, Magan F, Ahn S. The effect of rotational speed and the curvature of root canals on the breakage of rotary endodontic instruments. J Endod 2002;28(7):540–2.[Medline]

  31. Martin B, Zelada G, Varela P, et al. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J 2003;36(4): 262–6.[Medline]

  32. Janik JM. Access cavity preparation. Dent Clin North Am 1984;28(4):809–18.[Medline]

  33. Yared GM, Kulkarni GK. Failure of ProFile Ni-Ti instruments used by an inexperienced operator under access limitations. Int Endod J 2002;35(6):536–41.[Medline]

  34. Leeb J. Canal orifice enlargement as related to biomechanical preparation. J Endod 1983;9(11):463–70.[Medline]

  35. Morgan LF, Montgomery S. An evaluation of the crown-down pressureless technique. J Endod 1984;10(10):491–8.[Medline]

  36. Roland DD, Andeline WE, Browning DF, Hsu GH, Torabinejad M. The effect of preflaring on the rates of separation of 0.04 taper nickel titanium rotary instruments. J Endod 2002;28(7):543–5.[Medline]

  37. Patino PV, Biedma BM, Liebana CR, Cantatore G, Bahillo JG. The influence of a manual glide path on the separation rate of NiTi rotary instruments. J Endod 2005;31(2):114–6.[Medline]

  38. Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod 2004;30(4):228–30.[Medline]

  39. Yared GM, Bou Dagher, Kulkarni GK. Influence of torque control motors and operator’s proficiency on ProTaper file failures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003;96:229–33.[Medline]

  40. Gabel WP, Hoen M, Steiman HR, Pink FE, Dietz R. Effect of rotational speed on nickel-titanium file distortion. J Endod 1999;25(11):752–4.[Medline]

  41. Gambarini G. Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors. J Endod 2001;27(12):772–4.[Medline]

  42. Blum JY, Machtou P, Micallef JP. Location of contact areas on rotary Profile instruments in relationship to the forces developed during mechanical preparation on extracted teeth. Int Endod J 1999;32(2):108–14.[Medline]

  43. Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod 2000;26(3):161–5.[Medline]

  44. Bortnick KL, Steiman HR, Ruskin A. Comparison of nickel-titanium file distortion using electric and air-driven handpieces. J Endod 2001;27(1):57–9.[Medline]

  45. Yared GM, Bou Dagher FE, Machtou P. Failure of ProFile instruments used with high and low torque motors. Int Endod J 2001;34(6):471–5.[Medline]

  46. Yared G, Steiman P. Failure of ProFile instruments used with air, high torque control, and low torque control motors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2002;93(1):92–6.[Medline]

  47. Baker NA, Eleazer PD, Averbach RE, Seltzer S. Scanning electron microscopic study of the efficacy of various irrigating solutions. J Endod 1975;1(4):127–35.[Medline]

  48. McComb D, Smith DC. A preliminary scanning electron microscopic study of root canals after endodontic procedures. J Endod 1975;1(7):238–42.[Medline]

  49. Baumgartner JC, Mader CL. A scanning electron microscopic evaluation of four root canal irrigation regimens. J Endod 1987;13(4):147–57.[Medline]

  50. Stewart GG, Kapsimalas P, Rappaport H. EDTA and urea peroxide for root canal preparation. JADA 1969;78(2):335–8.

  51. Dederich DN, Zakariasen KL. The effects of cyclical axial motion on rotary endodontic instrument fatigue. Oral Surg Oral Med Oral Pathol 1986;61(2):192–6.[Medline]

  52. Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator’s proficiency on ProFile failure. Int Endod J 2001;34(1):47–53.[Medline]

  53. Yared GM, Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files. Int Endod J 2002;35(1):7–12.[Medline]

  54. Schrader C, Peters OA. Analysis of torque and force with differently tapered rotary endodontic instruments in vitro. J Endod 2005;31(2):120–3.[Medline]

  55. Di Fiore PM, Genov KI, Komaroff E, Dasanayake AP, Lin L. Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment. Int Endod J 2006;39(6):502–9.[Medline]

  56. Di Fiore PM, Genov KI, Komaroff E, Li Y, Lin LM. Nickel-titanium rotary instrument fracture: a clinical practice assessment. Int Endod J 2006;39(9):700–8.[Medline]





This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Di Fiore, P. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Di Fiore, P. M.
Related Collections
Right arrow Dental Equipment/Instruments


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS